Calibration
The calibration process reinitializes the persistent values that are device-specific. For this process, you need certified calibration weights. The app guides you through the entire process:
1. Unload the scale (ensure there is no weight on it) and perform the tare function by pressing and holding the button on the scale for more than 5 seconds (a blue LED will blink three times).
2. Place a calibration weight of a known value (for example, 1000 grams) on the scale, and enter this weight in the input below. Then, click ‘Run Calibration Process.’
The scale will calibrate the sensor and save the calibration values in the memory.
Successful calibration is indicated by a solid green LED blink. If an error occurs during weight measurement or calibration, the LED will blink red three times. In that case, repeat the process.